Troubleshooting Guide: H2O Labs Models 100E, 100SSE, 300E and 300SSE
Cooling fan does not operate or stops prematurely
The most common cause is the fan lead—the lead attached to the head or top of the unit—not being fully inserted into the socket on the body of the distiller.
The connection may appear secure even when the plug is not fully inserted. Push it firmly into the socket until the white line on the plug is no longer visible. The plug and socket are intentionally designed to fit snugly to reduce the possibility of water entering the connection.
If the fan continues running until the distillation cycle has finished, it is connected correctly.
Diagnostic check: To check whether the fan is working, connect the fan lead directly to the mains lead, bypassing the body of the distiller. Only perform this test as described in the instruction manual.
Water is very hot and steam is accumulating in the carafe
The cooling fan lead may not be fully inserted. Follow the checks in Cooling fan does not operate or stops prematurely.
The nozzle cannot be removed from the distiller head
Press the middle tabs on each side of the nozzle and tilt the nozzle out of its slots. Refer to the illustration in the instruction manual.
The nozzle may fit tightly. Hold the top of the unit firmly with one hand while carefully tilting and pulling the nozzle upwards and away from the unit with the other hand.
Alternatively, place the cooled and unplugged distiller head upside down on a stable, protected surface and carefully push the nozzle off.
The nozzle drops out of the distiller head
The nozzle is designed to hang from the outlet, so it may appear slightly loose. Make sure its tabs are seated and locked correctly.
It may help to bend the tabs out slightly before inserting the nozzle. If the tabs are worn and the nozzle continues to fall out, contact H2O Labs Technical Support about obtaining a replacement.
The glass insert cannot be removed from the nozzle
The glass insert that holds the filter pod can normally be cleaned without removing it from the nozzle.
When removal is necessary, gently press the top of the nozzle and place a small, flat-tipped screwdriver beneath the release tab while sliding the glass insert out. The instruction manual contains illustrations of this procedure.
It may help to have another person hold the release tab aside while you carefully pull out the glass insert. Take care to prevent the glass from breaking.
The fan and boiler are not working
Before assuming the boiler is not working: allow at least five minutes to determine whether it has started heating.
- Check that the mains socket works. Check the plug fuse, extension lead and consumer unit for a tripped circuit breaker or RCD. Only reset a breaker or RCD if it is safe to do so.
- If the mains supply is working, the appliance fuse or another internal component may need attention. Contact H2O Labs Technical Support or a qualified appliance repairer. Service charges may apply.
The fan works, but the boiler does not
- Allow approximately 20 minutes for the boiler to heat. Distilled water should then begin dripping into the carafe.
- If the boiler still does not heat after 20 minutes, unplug the distiller and contact H2O Labs Technical Support. The heating element or another component may require professional replacement.
The boiler works, but the fan does not
- Make sure the fan lead is pushed firmly and completely into the socket on the body of the distiller.
- Follow the diagnostic check under Cooling fan does not operate or stops prematurely.
- If the fan still does not operate, contact H2O Labs Technical Support. The distiller head assembly may need to be replaced.
The distiller turns off before completing a cycle
- Check for limescale in the bottom of the boiler. Limescale must be removed as described in the instruction manual. Otherwise, the thermal safety switch may shut down the unit to protect it from damage.
- Make sure the fan lead is pushed firmly and completely into its socket.
- If the distiller is clean and the fan lead is correctly connected, unplug the unit and contact H2O Labs Technical Support.
Steam escapes between the head and body
- Make sure the fan lead is pushed firmly and completely into its socket.
- The fan lead may be vibrating loose during a cycle. If the fan is not running when steam appears, switch off and unplug the distiller. Reconnect the fan lead firmly before using it again.
- If the silicone gasket was recently removed, check that it was reinstalled in the correct orientation.
- Dust on the fan or cooling coil may prevent steam from condensing efficiently. With the distiller unplugged and completely cool, clean the head as described in the instruction manual.
The distiller makes loud noises, bangs or vibrates excessively
Switch off and unplug the distiller. Excessive noise or vibration can occur when the fan is not operating and steam is not cooling and condensing correctly.
Make sure the fan lead is fully inserted and that the fan operates whenever the distiller is running. Do not continue using the distiller if excessive noise or vibration persists. Contact H2O Labs Technical Support.
The distiller produces little or no water
- Check whether air is exhausting from the top of the unit. If it is not, make sure the fan lead is fully inserted and the fan is operating.
- If the fan is working, allow approximately 20 minutes for the distiller to heat before water begins to collect.
- A full batch normally takes approximately five to six hours, depending on altitude, room temperature and source-water temperature.
- If the distiller remains unable to complete a cycle, unplug it and contact H2O Labs Technical Support.
The distiller starts before the reset button is pressed
The reset button may have been pressed during shipping or assembly. If so, the distiller will start when connected to the mains.
If you are not ready to begin a cycle, unplug the distiller. Connect it again when you are ready. The unit will switch off automatically at the end of the cycle. Press the reset button to begin subsequent cycles.
The distiller does not turn off when the reset button is pressed
This is normal. The reset button starts a cycle; it does not stop one. The thermal switch turns the unit off automatically when the cycle is complete.
To stop a cycle early, disconnect the distiller from the mains. Follow the instruction manual before reconnecting it and resuming operation.
The distiller does not turn on when the reset button is pressed
Follow the checks under The fan and boiler are not working.
The distiller runs only while the reset button is pressed
Unplug the distiller and contact H2O Labs Technical Support. The thermal switch or reset-button assembly may require professional replacement.
The reset-button frame is broken
Unplug the distiller and contact H2O Labs Technical Support about replacing the button assembly. A qualified appliance repairer should perform the repair. Service charges may apply.
Water is leaking onto the worktop
First identify where the leak originates. Do not touch or move a leaking distiller while it is connected to the mains.
- If the fan is not running, switch off and unplug the distiller. Once it is safe to do so, reconnect the fan lead firmly before starting another cycle.
- If water runs down the side of the carafe, reposition the activated carbon pod so it does not obstruct the opening at the bottom of the nozzle.
- If water leaks from inside the bottom of the distiller, unplug it immediately and stop using it. Contact H2O Labs Technical Support. Do not reconnect the appliance until it has been inspected.
The water has a slight smell or taste
New stainless-steel components can sometimes produce a slight new-appliance smell or taste during the first few cycles. Clean and prepare the distiller as directed in the instruction manual.
The smell or taste should fade after the first few cleaning and distillation cycles. If it persists, follow the guidance below.
The water continues to have an unpleasant smell or taste
The distiller vents VOCs through the condensing coil. The activated carbon pod provides additional reduction of residual smells and tastes.
In some areas, VOCs in the source water may evaporate near the end of a cycle and affect the smell or taste of the distilled water. The following steps may help:
- Stop the cycle approximately 30 to 60 minutes early by disconnecting the distiller from the mains. This leaves more source water in the boiling chamber, which must then be discarded. If using a plug-in timer, make sure it is appropriately rated for the appliance.
- Use a good-quality activated carbon filter to pre-filter the source water before distillation. Carbon filtration can help reduce VOCs that affect smell and taste.
If the problem continues, contact H2O Labs Technical Support.
The water has a higher-than-expected TDS or PPM reading
Many TDS meters require precise calibration and may be unreliable when testing distilled water. Check and calibrate the meter according to its manufacturer's instructions.
Small particles from the activated carbon pod may affect a TDS reading. To test without the pod, remove it and thoroughly wash the nozzle, glass insert and carafe before testing another batch.
If you are not using an activated carbon pod, a small amount of residue from cleaning or manufacturing may remain in the condensing coil. Running and discarding another cleaning cycle may resolve the issue. Follow the cleaning procedure in the instruction manual.
If readings remain unexpectedly high after confirming that the meter is correctly calibrated, contact H2O Labs Technical Support.